Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Philips PM 2718: (1 year) CAL VER /5520 DATE: 24-Sep-98 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 32 NUMBER OF LINES: 151 CONFIGURATION: Fluke 5520A ============================================================================= # # Compatibility: 5500/CAL or MET/CAL 5.0 or later. # # Subprocedures: None # # Required Files: None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R P J F W 1.002 ASK+ K X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Warm-up time: NONE 1.004 DISP Temperature : 23 degC +/-2 degC 1.004 DISP Humidity : 20% - 80% 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 HEAD {FREQUENCY} 1.007 DISP Connect the 5520A to the UUT as follows: 1.007 DISP [32] 5520A NORMAL HI to UUT V/OHM 1.007 DISP [32] 5520A NORMAL LO to UUT COM 1.008 DISP Press the button to turn the unit ON. 1.008 DISP 1.008 DISP Set the FUNCTION switch to Hz. 1.009 HEAD FREQUENCY: {20kHz Range} 1.010 5520 20 19.000kH 0.1% 0.01/ 0.7V SI 2W 2.001 HEAD FREQUENCY: {200kHz Range} 2.002 5520 200 199.00kH 0.1% 0.1/ 0.7V SI 2W 3.001 HEAD {LOGIC} 3.002 RSLT =Not Tested 3.003 JMP 4.002 3.004 DISP Press the button to select LOGIC. # 3.003 M5520 25pct 2.5Voff 3.005 5520 500H 5Vpp SQ S 2W 3.006 EVAL Check the LCD display for "_n__" 4.001 M5520 * 4.002 HEAD {AC VOLTAGE} 4.003 DISP Set the FUNCTION switch to AC VOLTS. 4.004 HEAD AC VOLTAGE: {1000mV Range} 4.005 5520 1000 1000.0mV 0.3% 0.1/ 1kH SI 2W 5.001 5520 1000 1000.0mV 5% 1.5/ 50kH SI 2W 6.001 HEAD AC VOLTAGE: {10V Range} 6.002 5520 10 10.000V 0.3% 0.1/ 1kH SI 2W 7.001 5520 10 10.000V 5% 1.5/ 50kH SI 2W 8.001 HEAD AC VOLTAGE: {100V Range} 8.002 5520 100 100.00V 0.3% 0.1/ 1kH SI 2W 9.001 5520 100 100.00V 2% 0.6/ 20kH SI 2W 10.001 HEAD AC VOLTAGE: {1000V Range} 10.002 5520 1000 900.0V 0.3% 0.1/ 100H SI 2W 11.001 HEAD {DC VOLTAGE} 11.002 DISP Set the FUNCTION switch to DC VOLTS. 11.003 HEAD DC VOLTAGE: {1000mV Range} 11.004 5520 1000 1000.0mV 0.04% 0.02/ 2W 12.001 HEAD DC VOLTAGE: {10V Range} 12.002 5520 10 10.000V 0.04% 0.02/ 2W 13.001 5520 10 -10.000V 0.04% 0.02/ 2W 14.001 HEAD DC VOLTAGE: {100V Range} 14.002 5520 100 100.00V 0.04% 0.02/ 2W 15.001 HEAD DC VOLTAGE: {1000V Range} 15.002 5520 1000 900.0V 0.1% 0.02/ 2W 16.001 HEAD {RESISTANCE} 16.002 DISP Set the FUNCTION switch to OHMS. 16.003 HEAD RESISTANCE: {100 MOhm Range} 16.004 5520 100 100.00MZ 5% 3/ 2W 17.001 HEAD RESISTANCE: {10 MOhm Range} 17.002 5520 10 10.000MZ 0.3% 0.1/ 2W 18.001 HEAD RESISTANCE: {1000 KOhm Range} 18.002 5520 1000 1000.0kZ 0.15% 0.05/ 2W 19.001 HEAD RESISTANCE: {100 KOhm Range} 19.002 DISP Connect the 5520A to the UUT as follows: 19.002 DISP [32] 5520A AUX HI to UUT V/OHM 19.002 DISP [32] 5520A AUX LO to UUT COM 19.002 DISP [32] 5520A NORMAL HI to UUT V/OHM 19.002 DISP [32] 5520A NORMAL LO to UUT COM 19.002 DISP Connect the SENSE leads DIRECTLY to the UUT. Connect 19.002 DISP the NORMAL leads to the SENSE leads AT the UUT. 19.003 5520 0.0Z S CW 19.004 DISP Press the button to remove the lead resistance. 19.005 5520 100 100.00kZ 0.15% 0.05/ CW 20.001 HEAD RESISTANCE: {10 KOhm Range} 20.002 5520 10 10.000kZ 0.15% 0.05/ CW 21.001 HEAD RESISTANCE: {1000 Ohm Range} 21.002 5520 1000 1000.0Z 0.15% 0.05/ CW 22.001 HEAD {CONTINUITY} 22.002 DISP Set the FUNCTION switch to CONTINUITY. 22.003 5520 1.0Z S CW 22.004 EVAL Does the UUT {display "00X.X" and an audible tone present}? 23.001 HEAD {DIODE} 23.002 DISP Press the button to select DIODE. 23.003 5520 1000 500mV 0.3% 0.1/ 2W 24.001 HEAD {DC CURRENT} 24.002 DISP Set the FUNCTION switch DC CURRENT. 24.003 DISP Connect the 5520A to the UUT as follows: 24.003 DISP [32] 5520A AUX HI to UUT mA 24.003 DISP [32] 5520A AUX LO to UUT COM 24.004 HEAD DC CURRENT: {20mA Range} 24.005 5520 20 20.00mA 0.4% 0.1/ 2W 25.001 HEAD DC CURRENT: {200mA Range} 25.002 5520 200 200.0mA 0.4% 0.1/ 2W 26.001 PIC pm2618hc 26.002 HEAD DC CURRENT: {2A Range} 26.003 5520 2 1.000A 0.4% 0.1/ 2W 27.001 HEAD DC CURRENT: {10A Range} 27.002 DISP Connect the 5520A to the UUT as follows: 27.002 DISP [32] 5520A 20A to UUT 10A 27.002 DISP [32] 5520A AUX LO to UUT COM 27.003 5520 10 10.00A 0.4% 0.1/ HC 2W 28.001 HEAD {AC CURRENT} 28.002 DISP Set the FUNCTION switch to AC CURRENT. 28.003 HEAD AC CURRENT: {10A Range} 28.004 5520 10 10.00A 0.4% 0.3/ 100H SI HC 2W 29.001 HEAD AC CURRENT: {2A Range} 29.002 DISP Connect the 5520A to the UUT as follows: 29.002 DISP [32] 5520A AUX HI to UUT mA 29.002 DISP [32] 5520A AUX LO to UUT COM 29.003 5520 2 1.000A 0.4% 0.3/ 100H SI 2W 30.001 HEAD AC CURRENT: {20mA Range} 30.002 5520 20 20.00mA 0.4% 0.3/ 100H SI 2W 31.001 HEAD AC CURRENT: {200mA Range} 31.002 5520 200 200.0mA 0.4% 0.3/ 100H SI 2W 32.001 END